Wrist-Based photoplethysmography assessment of heart rate and heart rate variability: Validation of WHOOP

Heart rate (HR) and HR variability (HRV) infer readiness to perform exercise in athletic populations. Technological advancements have facilitated HR and HRV quantification via photoplethysmography (PPG). This study evaluated the validity of WHOOP`s PPG-derived HR and HRV against electrocardiogram-derived (ECG) measures. HR and HRV were assessed via WHOOP and ECG over 15 opportunities. WHOOP-derived pulse-to-pulse (PP) intervals were edited with WHOOP`s proprietary filter, in addition to various filter strengths via Kubios HRV software. HR and HRV (Ln RMSSD) were quantified for each filter strength. Agreement was assessed via bias and limits of agreement (LOA), and contextualised using smallest worthwhile change (SWC) and coefficient of variation (CV). Regardless of filter strength, bias (=0.39 ± 0.38%) and LOA (=1.56%) in HR were lower than the CV (10-11%) and SWC (5-5.5%) for this parameter. For Ln RMSSD, bias (1.66 ± 1.80%) and LOA (±5.93%) were lowest for a 200 ms filter and WHOOP`s proprietary filter, which approached or exceeded the CV (3-13%) and SWC (1.5-6.5%) for this parameter. Acceptable agreement was found between WHOOP- and ECG-derived HR. Bias and LOA in Ln RMSSD approached or exceeded the SWC/CV for this variable and should be interpreted against its own level of bias precision.
© Copyright 2021 Sensors. All rights reserved.

Bibliographic Details
Subjects:
Notations:technical and natural sciences sports facilities and sports equipment
Tagging:Photoplethysmographie
Published in:Sensors
Language:English
Published: 2021
Online Access:https://doi.org/10.3390/s21103571
Volume:21
Issue:10
Pages:3571
Document types:article
Level:advanced