Comparison of heart-rate-variability recording with smartphone photoplethysmography, Polar H7 chest strap, and electrocardiography

(Vergleich der Aufzeichnung der Herzfrequenzvariabilität mittels Smartphone-Fotoplethysmografie, Polar H7-Brustgurt und EKG)

Purpose: To establish the validity of smartphone photoplethysmography (PPG) and heart-rate sensor in the measurement of heart-rate variability (HRV). Methods: 29 healthy subjects were measured at rest during 5 min of guided breathing and normal breathing using smartphone PPG, a heart-rate chest strap, and electrocardiography (ECG). The root mean sum of the squared differences between R-R intervals (rMSSD) was determined from each device. Results: Compared to ECG, the technical error of estimate (TEE) was acceptable for all conditions (average TEE CV% [90% CI]=6.35 [5.13; 8.5]). When assessed as a standardized difference, all differences were deemed "trivial" (average standard difference [90% CI]=0.10 [0.08; 0.13]). Both PPG- and heart-rate-sensor-derived measures had almost perfect correlations with ECG (R=1.00 [0.99; 1.00]). Conclusion: Both PPG and heart-rate sensors provide an acceptable agreement for the measurement of rMSSD when compared with ECG. Smartphone PPG technology may be a preferred method of HRV data collection for athletes due to its practicality and ease of use in the field.
© Copyright 2017 International Journal of Sports Physiology and Performance. Alle Rechte vorbehalten.

Bibliographische Detailangaben
Schlagworte:
Notationen:Trainingswissenschaft
Veröffentlicht in:International Journal of Sports Physiology and Performance
Sprache:Englisch
Veröffentlicht: 2017
Online-Zugang:https://doi.org/10.1123/ijspp.2016-0668
Jahrgang:12
Heft:10
Seiten:1324-1328
Dokumentenarten:Artikel
Level:hoch