Inter-machine reliability of the Biodex and Cybex isokinetic dynamometers for knee flexor/extensor isometric, concentric and eccentric tests
(Inter-Maschinen-Zuverlässigkeit der isokinetischen Dynamometer Biodex und Cybex für isometrische, konzentrische und exzentrische Kniebeuger-/Extensortests)
OBJECTIVES: To assess the inter-machine reliability of the Biodex System 3 Pro and Cybex Humac Norm Model 770 dynamometers for knee extensor and knee flexor peak torque measurements in isometric, concentric and eccentric tests.
DESIGN: Randomized/crossover.
SETTING: Exercise Research Laboratory, Federal University of Rio Grande do Sul (Brazil).
PARTICIPANTS: 25 healthy male subjects. MAIN OUTCOME MEASURES: Isometric, concentric and eccentric knee extensor and knee flexor peak torques recorded in the same test procedure performed on both isokinetic dynamometers. One-way ANOVA, intraclass correlation coefficient (ICC), standard error of measurement (SEM) and coefficient of variation (CV) were used to verify significant differences, relative and absolute reliability between devices.
RESULTS: No significant differences were found between tests performed on Biodex and Cybex (p > 0.05). ICC values indicated a high to very high reproducibility for isometric, concentric and eccentric peak torques (0.88-0.92), and moderate to high reliability for agonist-antagonist strength ratios (0.62-0.73). Peak torque did not show great difference between dynamometers for SEM (3.72-11.27 Nm) and CV (5.27-7.77%). Strength ratios presented CV values of 8.57-10.72%.
CONCLUSION Maximal knee extensor and knee flexor tests performed in isometric (60° of knee flexion), concentric and eccentric modes at 60°/s in Biodex and Cybex dynamometers present similar values.
© Copyright 2015 Physical Therapy in Sport. Elsevier. Veröffentlicht von Churchill Livingstone. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Trainingswissenschaft |
| Tagging: | isokinetisch |
| Veröffentlicht in: | Physical Therapy in Sport |
| Sprache: | Englisch |
| Veröffentlicht: |
Churchill Livingstone
2015
|
| Online-Zugang: | https://doi.org/10.1016/j.ptsp.2014.04.004 |
| Jahrgang: | 16 |
| Heft: | 1 |
| Seiten: | 59-65 |
| Dokumentenarten: | Artikel |
| Level: | hoch |