Validation of a new handheld device for measuring resting metabolic rate and oxygen consumption in children

(Validierung eines neuen Handheld-Gerätes zur Messung der Stoffwechselrate und des Sauerstoffverbrauchs von Kindern)

The purpose of this study was to assess the validity and reliability of the MedGem device to measure resting metabolic rate (RMR) in children. Subjects included 59 children (29 boys, 30 girls; mean age, 11.0 +/- 0.2 y). Subjects were given 4 RMR tests during 1 test session, cconsisting of 2 Douglas bag and 2 MedGem tests, in random counterbalanced order. No significant differences were found between Douglas bag and MedGem systems for oxygen consumption (209 +/- 5 and 213 +/- 5 mL/min, respectively, P = 0.106, r = 0.911, mean +/- standard deviation absolute difference 3.72 +/- 17.40 mL/min) or RMR (1460 +/- 39 and 1477 +/- 35 kcal/d, P = 0.286, r = 0.909, mean +/- standard deviation absolute difference 17.4 +/- 124 kcal/d). Standard error of estimates for oxygen consumption and RMR were 17.4 mL/min and 124 kcal/d, respectively. In conclusion, these data indicate that the MedGem is a reliable and valid system for measuring oxygen consumption and RMR in children.
© Copyright 2005 International Journal of Sport Nutrition and Exercise Metabolism. Human Kinetics. Alle Rechte vorbehalten.

Bibliographische Detailangaben
Schlagworte:
Notationen:Naturwissenschaften und Technik Biowissenschaften und Sportmedizin
Veröffentlicht in:International Journal of Sport Nutrition and Exercise Metabolism
Sprache:Englisch
Veröffentlicht: 2005
Online-Zugang:https://doi.org/10.1123/ijsnem.15.2.186
Jahrgang:15
Heft:2
Seiten:186-194
Dokumentenarten:Artikel
Level:hoch