Validity and reliability of the Valkyria Trainer Balance® portable force platform in its isometric mode
(Gültigkeit und Zuverlässigkeit der tragbaren Kraftplattform Valkyria Trainer Balance® in ihrem isometrischen Modus)
BACKGROUND: Isometric maximal voluntary force (IMVF) is essential for individuals` health and physical performance. Therefore, there is a need for valid and reliable devices to assess IMVF.
OBJECTIVE: To determine the validity and reliability of the Valkyria Trainer Balance® portable force platform in its isometric mode.
METHODS: Fifty-eight physically healthy individuals (30 men and 28 women) participated in the study. A repeated measures design was used to compare the inter-day test-retest reliability of peak force. The validity of the Valkyria Trainer Balance® force platform was determined by comparing the peak force with the ArtOficio® force platform. The analysis consisted of the intraclass correlation coefficient (ICC), standard error of measurement (SEM), and coefficient of variation (CV). A CV U+2A7D 10% and ICC U+2A7E 0.80 were considered acceptable reliability, while a U+2A7D 5% and ICC U+2A7E 0.90 were regarded as high reliability.
RESULTS: CV the peak force showed high test-retest inter-day reliability (CV = 4.3% and ICC = 0.99). When comparing both force platforms, there was a 1.1% difference between the two devices.
CONCLUSIONS: The results of this study demonstrate that the Valkyria Trainer Balance® force platform is valid and reliable for assessing IMVF in physically healthy individuals.
© Copyright 2024 Isokinetics and Exercise Science. IOS Press. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Naturwissenschaften und Technik |
| Tagging: | Kraftplattform Reliabilität Validität |
| Veröffentlicht in: | Isokinetics and Exercise Science |
| Sprache: | Englisch |
| Veröffentlicht: |
2024
|
| Online-Zugang: | http://doi.org/10.3233/IES-240022 |
| Jahrgang: | 32 |
| Heft: | 4 |
| Seiten: | 337-348 |
| Dokumentenarten: | Artikel |
| Level: | hoch |