Reliability and validity of the Lactate Pro 2 analyzer
(Reliabilität und Validität des Lactate Pro 2 Analyzers)
This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n=258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L-1 for blood lactate (BLa) =4.0 mmol.L-1 to -1.6 to +1.4mmol.L-1 for BLa >8.0 mmol.L-1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa = 8 mmol.L-1, and using the LP2 to identify training intensities at lactate threshold.
© Copyright 2021 Measurement in Physical Education and Exercise Science. Taylor & Francis. Veröffentlicht von Taylor & Francis. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Naturwissenschaften und Technik Biowissenschaften und Sportmedizin |
| Veröffentlicht in: | Measurement in Physical Education and Exercise Science |
| Sprache: | Englisch |
| Veröffentlicht: |
Philadelphia
Taylor & Francis
2021
|
| Online-Zugang: | https://doi.org/10.1080/1091367X.2020.1865966 |
| Jahrgang: | 25 |
| Heft: | 3 |
| Seiten: | 202-211 |
| Dokumentenarten: | Artikel |
| Level: | hoch |