Reliability and validity of the Lactate Pro 2 analyzer

(Reliabilität und Validität des Lactate Pro 2 Analyzers)

This study evaluated the reliability of the portable Lactate Pro 2 analyzer (LP2), and its validity compared to a laboratory-based analyzer, YSI 1500 Sport (YSI). Blood samples (n=258) were collected during 44 graded incremental rowing tests, with data from 17 tests used to quantify load, heart rate, and oxygen consumption at lactate threshold. Inter-LP2 reliability was high with coefficient of variation = 3.3%; Bland Altman 95% limits of agreement varied from ± 0.3 mmol.L-1 for blood lactate (BLa) =4.0 mmol.L-1 to -1.6 to +1.4mmol.L-1 for BLa >8.0 mmol.L-1, with minimal systematic bias. Despite LP2 measurements being ~1.42 times higher than YSI measurements, no significant differences (P > .05) were detected when quantifying load, heart rate, or oxygen consumption data at lactate threshold. The results support using different LP2 devices interchangeably, especially when BLa = 8 mmol.L-1, and using the LP2 to identify training intensities at lactate threshold.