Validation of the photoelectric optogait system to measure racewalking biomechanical parameters on a treadmill

(Validierung eines photoelektrischen Optogait-Systems zur Messung biomechanischer Parameter beim Gehen auf dem Laufband)

The purpose of this pilot study was to validate the Optogait system to measure racewalking biomechanic parameters on a treadmill. Contact time, flight time, cadence and step length are some of the biomechanic parameters which influence the performance in racewalking. Five subjects were analyzed while racewalking at different speeds on a treadmill. The Optogait system was compared with a high speed video used as reference system. Flight time was overestimated by Optogait when compared with the reference system (0.025 ± 0.014 vs 0.023 ± 0.014 seconds, t = -2.43, p <0,05). No differences were found in contact time or step length. Thus, the present study validates the Optogait system to measure racewalking biomechanical parameters on a treadmill
© Copyright 2017 ISBS Proceedings Archive (Michigan). Northern Michigan University. Veröffentlicht von International Society of Biomechanics in Sports. Alle Rechte vorbehalten.

Bibliographische Detailangaben
Schlagworte:
Notationen:Trainingswissenschaft Naturwissenschaften und Technik Ausdauersportarten
Veröffentlicht in:ISBS Proceedings Archive (Michigan)
Sprache:Englisch
Veröffentlicht: Cologne International Society of Biomechanics in Sports 2017
Online-Zugang:https://commons.nmu.edu/isbs/vol35/iss1/253
Jahrgang:35
Heft:1
Seiten:292-294
Dokumentenarten:Kongressband, Tagungsbericht
Level:hoch