Validation of the photoelectric optogait system to measure racewalking biomechanical parameters on a treadmill
(Validierung eines photoelektrischen Optogait-Systems zur Messung biomechanischer Parameter beim Gehen auf dem Laufband)
The purpose of this pilot study was to validate the Optogait system to measure racewalking biomechanic parameters on a treadmill. Contact time, flight time, cadence and step length are some of the biomechanic parameters which influence the performance in racewalking. Five subjects were analyzed while racewalking at different speeds on a treadmill. The Optogait system was compared with a high speed video used as reference system. Flight time was overestimated by Optogait when compared with the reference system (0.025 ± 0.014 vs 0.023 ± 0.014 seconds, t = -2.43, p <0,05). No differences were found in contact time or step length. Thus, the present study validates the Optogait system to measure racewalking biomechanical parameters on a treadmill
© Copyright 2017 ISBS Proceedings Archive (Michigan). Northern Michigan University. Veröffentlicht von International Society of Biomechanics in Sports. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Trainingswissenschaft Naturwissenschaften und Technik Ausdauersportarten |
| Veröffentlicht in: | ISBS Proceedings Archive (Michigan) |
| Sprache: | Englisch |
| Veröffentlicht: |
Cologne
International Society of Biomechanics in Sports
2017
|
| Online-Zugang: | https://commons.nmu.edu/isbs/vol35/iss1/253 |
| Jahrgang: | 35 |
| Heft: | 1 |
| Seiten: | 292-294 |
| Dokumentenarten: | Kongressband, Tagungsbericht |
| Level: | hoch |