Relative body weight and standardised brightness-mode ultrasound measurement of subcutaneous fat in athletes: An international multicentre reliability study, under the auspices of the IOC Medical Commission
(Relatives Körpergewicht und standardisierte Helligkeitsmodus-Ultraschallmessung des subkutanen Fetts bei Sportlern: Eine internationale Zuverlässigkeitsstudie mit Beteiligung mehrerer Zentren unter der Schirmherrschaft der Mdizinischen Kommission des IOC)
Introduction: Fat is a metabolic fuel, but excess body fat is ballast mass, and therefore, many elite athletes reduce body fat to dangerously low levels. Uncompressed subcutaneous adipose tissue (SAT) thickness measured by brightness-mode ultrasound (US) provides an estimate of body fat content.
Methods: The accuracy for determining tissue borders is about 0.1-0.2 mm and reliability (experienced measurers) was within ± 1.4 mm (95% limit of agreement, LOA). We present here inter- and intra-measurer scores of three experienced US measurers from each of the centres C1 and C2, and of three novice measurers from each of the centres C3-C5. Each of the five centres measured 16 competitive adult athletes of national or international level, except for one centre where the number was 12. The following sports were included: artistic gymnastics, judo, pentathlon, power lifting, rowing, kayak, soccer, tennis, rugby, basketball, field hockey, water polo, volleyball, American football, triathlon, swimming, cycling, long-distance running, mid-distance running, hurdles, cross-country skiing, snowboarding, and ice hockey. SAT contour was detected semi-automatically: typically, 100 thicknesses of SAT at a given site (i.e., in a given image), with and without fibrous structures, were measured.
Results: At SAT thickness sums DI (of eight standardised sites) between 6.0 and 70.0 mm, the LOA of experienced measurers was 1.2 mm, and the intra-class correlation coefficient ICC was 0.998; novice measurers: 3.1 mm and 0.988. Intra-measurer differences were similar. The median DI value of all 39 female participants was 51 mm (11% fibrous structures) compared to 17 mm (18%) in the 37 male participants.
Discussion: DI measurement accuracy and precision enables detection of fat mass changes of approximately 0.2 kg. Such reliability has not been reached with any other method. Although females` median body mass index and mass index were lower than those of males, females` median DI was three times higher, and their percentage of fibrous structures was lower. The standardised US method provides a highly accurate and reliable tool for measuring SAT and thus changes in body fat, but training of measurers is important.
© Copyright 2020 Sports Medicine. Springer. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Biowissenschaften und Sportmedizin |
| Veröffentlicht in: | Sports Medicine |
| Sprache: | Englisch |
| Veröffentlicht: |
2020
|
| Online-Zugang: | https://doi.org/10.1007/s40279-019-01192-9 |
| Jahrgang: | 50 |
| Heft: | 3 |
| Seiten: | 597-614 |
| Dokumentenarten: | Artikel |
| Level: | hoch |