Measurement accuracy of head impact monitoring sensor in sport

Head injury and brain trauma exposure in sport have been recognized as potential contributors to long-term neurological disorders. As a result sensors have been proposed as an impact severity monitoring tool for on-field measurement of head accelerations. The purpose of this study was to characterize the accuracy of a head impact monitoring sensor system. Peak acceleration responses from a Smart lmpact Monitor (SIM) sensor were compared against reference sensors from a Hodgson-WSU headform. The headform with SIM was impacted for 7 impact conditions and 3 inbound energies. Moderate to strong positive correlations were found between the SIM and reference sensors for all impact conditions. At higher inbound energy the SIM overestimated, suggesting that under higher risk conditions the SIM represents a conservative tool in identifying dangerous impacts.
© Copyright 2016 ISBS - Conference Proceedings Archive (Konstanz). Springer. Published by International Society of Biomechanics in Sports. All rights reserved.

Bibliographic Details
Subjects:
Notations:biological and medical sciences technical and natural sciences
Tagging:Kopfschutz
Published in:ISBS - Conference Proceedings Archive (Konstanz)
Language:English
Published: Tsukuba International Society of Biomechanics in Sports 2016
Online Access:https://ojs.ub.uni-konstanz.de/cpa/article/view/6994
Volume:34
Issue:1
Pages:51-54
Document types:congress proceedings
Level:advanced