Stiffness comparison of running prosthetic feet of different categories after bench testing
(Vergleich der Steifigkeit von Fußprothesen zum Laufen verschiedener Kategorien nach dem Prüfstandtest)
The purpose of the work is the comparison of stiffness properties of Running Prosthetic Feet (RPF) for transtibial amputee athletes after the introduction of test methods to collect and analyse their Load-deflection curves. The study explores the effect of the orientation of the socket (Theta G) with respect to ground during a load cycle.The three Ossur feet Cheetah Xtreme Category 4th, 5th ,6th underwent extensive testing on a multi-component test bench. Results show that the unit interval between categories does not matches with the interval in terms of equivalent stiffness Keq introduced as synthetic stiffness parameter.
© Copyright 2022 ISBS Proceedings Archive (Michigan). Northern Michigan University. Veröffentlicht von International Society of Biomechanics in Sports. Alle Rechte vorbehalten.
| Schlagworte: | |
|---|---|
| Notationen: | Parasport Naturwissenschaften und Technik |
| Tagging: | Prothese Steifigkeit Dämpfung Amputierte |
| Veröffentlicht in: | ISBS Proceedings Archive (Michigan) |
| Sprache: | Englisch |
| Veröffentlicht: |
Liverpool
International Society of Biomechanics in Sports
2022
|
| Online-Zugang: | https://commons.nmu.edu/isbs/vol40/iss1/136/ |
| Jahrgang: | 40 |
| Heft: | 1 |
| Seiten: | Article 136 |
| Dokumentenarten: | Kongressband, Tagungsbericht |
| Level: | hoch |